Description
Optical Technology without calibration
Oxygen Optical Digital Sensor

Oxygen Optical Digital Sensor 7m OPTOD for ODEON
Fischer connector for ODEON (handheld multiparameter device) and Module 4200 (USB – RS485 converter).
The OPTOD (Optical Dissolved Oxygen technology) is based on luminescent optical technology. The OPTOD sensor is approved by the ASTM International Method D888-05 and by the International norm ISO 17289.
Without calibration requirements and thanks to an ultra-low power technology, the OPTOD sensor meets the demands of field works and short or long term campaigns. Without oxygen consumption, this technology allows you an accurate measure in all situations and especially in very low oxygen concentrations
Advantages
- Optical Technology without calibration
- Digital Technology (Modbus RS-485)
- No drift, Reduced maintenance
- Body in Stainless steel (316 L) or Titanium
Optical Technology
The OPTOD (Optical Dissolved Oxygen technology) is based on luminescent optical technology. The OPTOD sensor is approved by the ASTM International Method D888-05.
Without calibration requirements and thanks to an ultra low power technology, the OPTOD sensor meets the demands of field works and short or long term campaigns.
Without oxygen consumption, this technology allows you an accurate measure in all situation and especially in very low oxygen concentrations
Digital Technology
The “smart” OPTOD sensor stores calibration and history data within the sensor. This allows you a“plug and play”system without re-calibration.
Thanks to the Universal Modbus RS485 protocol, the PONSEL OPTOD can be connected to all devices commonly used (Datalogger, Controller, Automat, Remote System…).
Mechanic
Compact, strong and light, the sensor allows a portable or in fixed/permanent use.
Body in Stainless steel 316 L (passivation treatment) or in Titanium for applications in corrosive environment.
- Urban wastewater treatment
- Industrial effluent treatment
- Surface water monitoring,
- Sea water monitoring
- Fish farming
- Aquarium
- Drinking water
Measures
Measure Principles | Optical measure by luminescence |
---|---|
Measure ranges | 0,00 to 20,00 mg/L 0,00 to 20,00 ppm 0-200% |
Resolution | 0,01 |
Accuracy | +/- 0,1mg/L +/- 0,1 ppm +/-1% |
Response time | 90% of the value in less than 60 seconds |
Frequency of recommended measure | >5s |
Resolution | 0,01 |
Water move | Via NTC |
Stocking temperature | -10°C to + 60°C |
Signal interface | Modbus RS-485 (standard) and SDI-12 (option) |
Sensor power-supply | 5 to 12 volts |
Consumption | Standby 25 μA Average RS485 (1 measure/ seconde) : 4,4 mA Average SDI12 (1 measure/ seconde) : 7,3 mA Current pulse : 100 mA |
Sensor
Dimensions | Diameter : 25 mm ; length : 146 mm |
---|---|
Weight | Stainless steel version 450g (sensor + cable 3 m) Titanium version 300 g (sensor + cable 3 m) |
Material | Stainless steel 316L, New : body in Titanium |
Maximum pressure | 5 bars |
Connection | 9 armored connectors, polyurethane jacket, bare wires or waterproof Fisher connector |
Protection | IP68 |